Patient stopping-power accuracy below one percent · revision history
Revision 1
Accepted · first revision · record
- Title
- Patient stopping-power accuracy below one percent
- Plain
- The plan converts a CT scan into a map of how fast particles slow down. Converting image numbers into that map introduces an error of a few percent, which becomes a few millimetres of range error.
- Precise
- Determine the stopping power of patient tissues relative to water from imaging to better than 1 percent root-mean-square, including tissue mean excitation energies relative to water and tissue-composition variability. Candidate routes: dual-energy CT, photon-counting spectral CT, proton CT. An answer is a validated conversion with a stated uncertainty budget.
- Settled by
- Direct proton-beam measurement of stopping power in tissue samples and phantoms compared against the imaging-derived map.
- Kind
- well-posed
- Answer type
- value
- References
- (empty)
- Literature status
- partially-resolved
- Status note
- Dual-energy CT lowered relative stopping-power errors to roughly 1 percent by 2023; tissue I-values and tissue-composition variability remain the dominant residual terms.
- Posed since
- 1996