Sensor material choice for clinical count rates · revision history

Revision 1

Accepted · first revision · record

Title
Sensor material choice for clinical count rates
Plain
The detector can be made of cadmium telluride, cadmium zinc telluride or silicon, each failing in a different way at high photon rates. Which is best overall is unsettled.
Precise
Compare cadmium telluride, cadmium zinc telluride and deep-silicon sensors on detection efficiency at 20 to 140 keV, charge-transport polarization at fluxes above $10^{8}$ counts per second per $\mathrm{mm}^{2}$, charge-sharing fraction at a given pixel pitch, and K-escape, under identical clinical conditions. An answer is a classification of which material wins for which imaging task, with the physical reasons.
Settled by
Side-by-side measurement of the same imaging tasks on detectors of each material at clinical flux.
Kind
well-posed
Answer type
classification
References
(empty)
Literature status
contested
Status note
CdTe-based clinical systems have been in use since 2021; deep-silicon systems moved from clinical prototypes toward commercial release around 2025, and no side-by-side comparison under identical conditions has settled the question.
Posed since
2012