Is 10 ps coincidence time resolution reachable? · revision history

Revision 1

Superseded · first revision · record

Title
Is 10 ps coincidence time resolution reachable
Plain
Today the best clinical scanners time the two photons to about two hundred trillionths of a second (picoseconds). Getting to ten would place each emission directly, with no image reconstruction.
Precise
Determine whether a coincidence time resolution of 10 ps full width at half maximum is attainable in a detector of sufficient stopping power for 511 keV photons, which corresponds to a position uncertainty of about 1.5 mm along the line of response. Required chain: photon production statistics and rise time, light transport, photodetector transit-time spread, and electronics jitter. An answer is a demonstrated detector pair or a proof that a chain element forbids it.
Settled by
A bench measurement of 10 ps coincidence resolution between two detectors with 511 keV efficiency comparable to clinical crystals.
Kind
well-posed
Answer type
yes-no
References
(empty)
Literature status
open
Status note
Laboratory detector pairs reached the tens of ps range by the mid 2020s; clinical systems were at about 180 to 230 ps.
Posed since
2017