Quantitative imaging of low-yield therapeutic emitters · revision history
Revision 1
Accepted · first revision · record
- Title
- Quantitative imaging of low-yield therapeutic emitters
- Plain
- To compute the dose, the drug must be seen and counted inside the body, but alpha-emitting drugs give off very few detectable photons.
- Precise
- Achieve activity quantification to better than 10 percent for therapeutic nuclides with sparse or complex photon emission, notably Ac-225 at administered activities of order 2.7 microCi per kg (about 0.2 mCi per patient), imaged through the daughter gamma lines at 218 keV (Fr-221) and 440 keV (Bi-213), and Pb-212 at 239 keV, including scatter, collimator septal penetration and partial-volume correction. An answer is a calibrated imaging and reconstruction protocol with measured accuracy.
- Settled by
- Phantom and patient measurements with independent activity ground truth demonstrating the stated accuracy.
- Kind
- well-posed
- Answer type
- value
- References
- (empty)
- Literature status
- open
- Status note
- (empty)
- Posed since
- 2015